1.Product introduction
G01NET-3(d) high-precision data acquisition instrument is a dynamic data acquisition instrument with LCD touch screen, which has the functions of monitoring data synchronous acquisition, field storage and real-time data upload. The acquisition instrument can be calibrated through the upper computer network; the network port or serial port can be selected to transmit raw data or analysis results. Collector built-in 16 G data memory card. Data in the memory card can be copied through the USB port of the instrument.
This instrument supports customers` secondary development. We can open the communication protocol of the acquisition instrument free of charge.
Please read the software instructions for detailed use of the software, such as the instrument supporting special parameter setting software, data acquisition and analysis software(01NETCS2019S1 Data acquisition instrument Parameter setting software Instructions、Operation Manual of G01NETFDC2020S1 Wind power tower safety monitoring system、Operation Manual of DZJC2020S1 Ground motion and Vibration monitoring system)
2.Main technical parameters of the product
Name of Index | Technical Parameter |
Type | G01NET-3D |
Number of Input Channels | 8~16 Channel(Default single-ended input, support differential input) |
Input Range and Resolution | -10~10V/0.005mV |
Sampling Freqency | 16~600Hz(All channels are fully synchronized) |
Storage Capacity and Format | Storage capacity:standard configuration 16G SD card,Storage format:Text format |
Remote Data Transimission
| It supports UDP, TCPIP Client and TCPIP Server. Default is TCPIP Server; Continuous transmission is the default after power on. 4GDTU can be used for data transmission |
Time Correction Function | Manual calibration, remote server calibration |
Display Function | 7 inch or 4.3 inch LCD, |
Dynamic Range | ≥120dB |
Power Supply | DC12V |
Power Dissipation | 7 watts |
Size(L*W*H,mm) | 270*200*120 |
Input impedence | 1M Ohm |
External Interface | 7 inch LCD screen, 1 serial port,2 USB ports, 1 Internet port 4.3 inch LCD screen : 1 serial,2 USB ports,1 Internet port
|
Operatiing Temperature | Temperature:-35℃~+70℃ :Humidity≤90% |
Trriger Cut-off Frenquency | Soft trigger, default 20Hz, settable |
Field Data Storage Trigger mode | Trigger based on amplitude
|
3.Application scope
Blasting monitoring, fan tower, bridge and other buildings safety monitoring
4.Product features
High precision, low power consumption, small volume, real-time data acquisition, triggered data storage, real-time data transmission, man-machine operation interface.
5.The Introduction of Instrument Appearance
No. | Name | Desicription |
① | COM | Data transmission interface requires serial port to network port module to be used, standby function. |
② | 12V output | Mark (+) on the connection terminal, connect the sensor's power cord to supply power to the sensor, multiple sensors can be shared. |
③ | Ground Wire(GND) | Mark (-) on the terminal, connect the sensor ground wire and shield wire and temperature compensation wire, multiple sensors can be shared. |
④ | Signal Channel | The labels 1~16 on the wiring terminals belong to the signal channel, which is connected to the sensor signal line. Only one signal line is allowed for each channel. |
⑤ | USB | Use USB interface to connect mouse or U disk |
⑥ | Scerrn | Display the operation status and parameter setting of the acquisition instrument |
⑦ | 12V input | Terminal block type 12V DC input interface |
⑧ | 12V input | The jack type 12V DC input interface supplies power to the collector, and the
power adapter is standard at the factory. |
⑨ | LAN | Universal network port, data transmission interface, connect to switch or server. |
⑩ | Reset Button | Power on for 5 seconds, restore the factory Settings of the network port IP and port number of the acquisition instrument |
Terminal block detailed description:There are 22 interfaces in the terminal block of the collector, counting from right to left. Interfaces 1, 2, and 3 are the positive poles of the power supply for the sensor (marked with "+" on the label), 4, 5, and 6 are the grounds for connecting the sensor. The ground wire of the power supply of the wire and the sensor (marked "-" on the label). Ports 7 to 22 are labeled "1"-"16" on the label of the analog signal input interface, and are connected to the sensor signal line.Please connect in this order if the labels and ports do not correspond one by one. |